Appendix G. CR10X Programming
Example
Although this example is for the CR10X, the CR23X is programmed similarly.
;{CR10X}
;
;
;
*Table 1 Program
01: 900
Execution Interval (seconds)
1: Do (P86)
1: 42
Set Port 2 High
2: Beginning of Loop (P87)
1: 0000
Delay
2: 16
Loop Count
3: Do (P86)
1: 73
Pulse Port 3
4: Excitation with Delay (P22)
1: 1
Ex Channel
2: 0000
Delay W/Ex (0.01 sec units)
3: 10
Delay After Ex (0.01 sec units)
4: 0000
mV Excitation
5: SDI-12 Recorder (P105)
1: 0
SDI-12 Address
2: 1
Start Measurement (aM1!)
3: 1
Port
4: 91
Loc [ VW ]
5: 1.0
Multiplier
6: 0.0
Offset
6: Z=X (P31)
;Write Freq to inlocs 1-16
1: 91
X Loc [ VW ]
2: 1
-- Z Loc [ Freq_1 ]
7: Z=X (P31)
;Write Amp to inlocs 17-32
1: 92
X Loc [ Amplitude ]
2: 17
-- Z Loc [ Amp_1 ]
8: Z=X (P31)
;Write Signal to Noise ratio to inlocs 33-48
1: 93
X Loc [ Sig2Noise ]
2: 3
-- Z Loc [ S2Noise_1 ]
9: Z=X (P31)
;Write Frequency of Noise to inlocs 49-64
1: 94
X Loc [ NoiseFreq ]
2: 49
-- Z Loc [ NsFreq_1 ]
G-1