Campbell Scientific AVW200-series Vibrating Wire Interfaces User Manual
Page 81

Appendix D. The Public Table
Control Parameters When Troubleshooter is Running
TimeSeries(1) Writing this variable will force a Vibrating Wire
measurement and create the TimeSeries.bin and
spectrum.bin files.
Example:
101 = measures AVW200 chan1 and Mux chan1
102 = measures AVW200 chan1 and Mux chan2
.....
201 = measures AVW200 chan2 and Mux chan1
202 = measures AVW200 chan2 and Mux chan2
....
232 = measures AVW200 chan2 and Mux chan32
Notes: TimeSeries(2)..(4) need to be filled with valid
values before making the measurement.
TimeSeries(2) Begin Frequency for force measurement (100 - 6500)
TimeSeries(3) Eng Frequency for forcing measurement (100 - 6500)
TimeSeries(4) Excite voltage for forcing measurement (1=5volts or
2=12volts)
TimeSeries(5) Copy of the Chan/MuxChan executed in TimeSeries(1)
TimeSeries(6) Measured Frequency (Hz)
TimeSeries(7) Measured Amplitude (MVolts_RMS)
TimeSeries(8) Measured Signal-to-Noise Ratio
TimeSeries(9) Measured Noise Freq (Hz)
TimeSeries(10) Measured Decay Ratio
TimeSeries(11) Measured Thermistor (ohms)
D-3