Campbell Scientific CR510 Basic Datalogger User Manual

Page 152

Advertising
background image

SECTION 13. CR510 MEASUREMENTS

13-8

TABLE 13.3-4. Measured Peak Excitation Transients for 1000 Foot Lengths of Three Belden Lead

Wires Used by Campbell Scientific

-----------------------V

eo

(mV) -----------------------

Vx(mV)

Rf=1 kohm

R

f

=10 kohm

#

#

#

#

#

#

8641

8771

8723

8641

8771

8723

2000

50

100

60

100

140

80

1000

25

65

40

60

90

40

NOTE: Excitation transients are eliminated
if excitation leads are contained in a shield
independent from the signal leads.

The size of the peak transient is linearly related
to the excitation voltage and increases as the
bridge resistor, R

f

, increases. Table 13.3-4

shows measured levels of V

eo

for 1000 foot

lengths of three Belden wires used in Campbell
Scientific sensors. Values are given for R

f

equal to 1 kohm and 10 kohm. Table 13.3-4 is
meant only to provide estimates of the size of
excitation transients encountered; the exact
level will depend upon the specific sensor
configuration.

Equation 13.3-7 can be solved for the maximum
lead length, L, permitted to maintain a specified
error limit. Combining Equations 13.3-7 and
13.3-4 and solving for L gives:

L = -(R

o

C

f

+ (t/ln(V

e

/V

eo

)))/R

o

C

w

[13.3-15]

where V

e

is the measurement error limit.

EXAMPLE LEAD LENGTH CALCULATION
FOR 107 TEMPERATURE SENSOR

Assume a limit of 0.05

°

C over a 0

°

C to +40

°

C

range is established for the transient settling
error. This limit is a reasonable choice since it
approximates the linearization error over that
range. The output signal from the thermistor
bridge varies nonlinearly with temperature
ranging from about 100 µV/

°

C at 0

°

C to 50

µV/

°

C at 40

°

C. Taking the most conservative

figure yields an error limit of V

e

= 2.5 µV. The

other values needed to calculate the maximum
lead length are summarized in Table 13.3-5 and
listed below:

1) V

eo

~

50 mV, peak transient at 2 V excitation

2) V

e

~ 2.5 µV, allowable measurement error

3) t = 450 µs, CR510 input settling time

4) R

o

= 1 kohm, 107 probe source resistance

5) C

f

= 3.3 nfd, CR510 input capacitance

6) C

w

~

42 pfd/ft., lead wire capacitance

Solving Equation 13.3-15 gives a maximum
lead length of:

L

~

1003 ft., error

~

0.05

°

C

Setting the allowable error at 0.1

°

C or

approximately 5 µV, the maximum lead length
increases to:

L

~

1085 ft., error

~

0.1

°

C

13.3.4 SUMMARY OF SETTLING ERRORS FOR

CAMPBELL SCIENTIFIC RESISTIVE
SENSORS

Table 13.3-5 summarizes the data required to
estimate the effect of lead length on settling
errors for Campbell Scientific's resistive
sensors. Comparing the transient level, V

eo

, to

the input range, one suspects that transient
errors are the most likely limitation for the 107
sensor. The sensors in the WVU-7 are the
same as in the Model 107 (the lead wire is
different), but the signal leads for the WVU-7
wet- and dry-bulbs are not subject to excitation
transients because they are shielded
independently from the excitation.

Advertising