5 voltage range and overrange detection, 6 output processing – Campbell Scientific CR510 Basic Datalogger User Manual

Page 50

Advertising
background image

SECTION 3. INSTRUCTION SET BASICS

3-2

Loop Index, allows the increment step to be
changed. See Instructions 87 and 90, Section
12, for more details.

To index an input location (4 digit integer), C or
"-" is pressed after keying the value but before
entering the parameter. Two minus signs (--)
will be displayed to the right of the parameter.

3.5 VOLTAGE RANGE AND

OVERRANGE DETECTION

The voltage RANGE code parameter on
Input/Output Instructions is used to specify the
full scale range of the measurement and the
integration period for the measurement (Table
3.5-1).

The full scale range selected should be the
smallest that will accommodate the full scale
output of the sensor being measured. Using
the smallest possible range will result in the
best resolution for the measurement.

Four different integration sequences are
possible. The relative immunity of the
integration sequences to random noise is: 60
Hz rej. = 50 Hz rej. > 2.72ms integ. > 272

µ

s

integ. The 60 Hz rejection integration rejects
noise from 60 Hz AC line power. The 50 Hz
rejection is for countries whose electric utilities
operate at 50 Hz (Section 13.1).

When a voltage input exceeds the range
programmed, the value which is stored is set to
the maximum negative number and displayed
as -99999 in high resolution or -6999 in low
resolution.

An input voltage greater than +5 volts on one of
the analog inputs will result in errors and
possible overranging on the other analog inputs.

Voltages greater than 16 volts may permanently
damage the CR510.

NOTE: Voltages in excess of 5.5 volts
applied to a control port can cause the
CR510 to malfunction.

3.6 OUTPUT PROCESSING

Most Output Processing Instructions have both
an Intermediate Data Processing operation and
a Final Data Processing operation. For
example, when the Average Instruction, 71, is
initiated, the intermediate processing operation
increments a sample count and adds each new
Input Storage value to a cumulative total
residing in Intermediate Storage. When the
Output Flag is set, the final processing
operation divides the cumulative total by the
number of samples to find the average. The
average is then stored in final storage and the
cumulative total and number of samples are set
to zero in Intermediate Storage.

Final Storage Area 1 (Sections 1.5, 2.1) is the
default destination of data output by Output
Processing Instructions. Instruction 80 may be
used to direct output to either Final Storage
Area 2 or to Input Storage.

Output Processing Instructions requiring
intermediate processing sample the specified
input location(s) each time the Output
Instruction is executed, NOT each time the
location value is updated by an I/O Instruction.
For example: Suppose a temperature
measurement is initiated by Table 1 which has
an execution interval of 1 second.

TABLE 3.5-1. Input Voltage Ranges and Codes

Range Code

Full Scale Range

Resolution*

Slow

Fast

2.72ms

250 us

60 Hz

50 Hz

Integ.

Integ. Reject.

Reject.

1

11

21

31

±

2.5 mV

0.33 µV

2

12

22

32

±

7.5 mV

1.0

µV

3

13

23

33

±

25

mV

3.33 µV

4

14

24

34

±

250

mV

33.3

µV

5

15

25

35

±

2500

mV

333.

µV

* Differential measurement, resolution for single-ended measurement is twice value shown.

Advertising