SECTION 2. INTERNAL DATA STORAGE

2-4

another memory location may be entered,
followed by the "A" key to jump to the start of
the Output Array equal to or just ahead of the
location entered. Whenever a location number
is displayed by using the "#" key, the
corresponding data point can be displayed by
pressing the "C" key.

The same element in the next Output Array with
the same ID can be displayed by hitting #A.
The same element in the previous array can be
displayed by hitting #B. If the element is 1
(Array ID), then #A advances to the next array
and #B backs up to the previous array. #0A
backs up to the start of the current array.

The keyboard commands used in the

7 Mode

are summarized in Table 2.3-1.

Advancing the DPTR past the Data Storage
Pointer (DSP) displays the oldest data point.
Upon entering the

7 Mode, the oldest Output

Array can be accessed by pressing the "A" key.

TABLE 2.3-1.

∗∗∗∗

7 Mode Command Summary

Key

Action

A

Advance to next data point

B

Back-up to previous data point

#

Display location number of currently
displayed data point value

C

Display value of current location

#

A

Advance to same element in next
Output Array with same ID

#

B

Back-up to same element in
previous Output Array with same ID

#

0

A

Back-up to the start of the current
Final Data Storage Array

Exit

7 Mode