SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES
7-16
CR510
12 V or 5 V
Ch E1
FIGURE 7.13-3. Hook up to AVW1
PROGRAM
AVW1 & CR510 USED TO MEASURE 1
GEOKON VIBRATING WIRE SENSOR.
* Table 1 Program
01:
60
Execution Interval (seconds)
01:
Excite-Delay (SE) (P4)
1:
1
Reps
2:
15
±
2500 mV Fast Range
3:
1
SE Channel
4:
1
Excite all reps w/Exchan 1
5:
1
Delay (units 0.01 sec)
6: 2500
mV Excitation
7:
1
Loc [ Temp ]
8:
.001
Mult
9:
0
Offset
02:
Polynomial (P55)
1:
1
Reps
2:
1
X Loc [ Temp ]
3:
1
F(X) Loc [ Temp ]
4: -104.78
C0
5: 378.11
C1
6: -611.59
C2
7: 544.27
C3
8: -240.91
C4
9:
43.089
C5
03:
Vibrating Wire (SE) (P28)
1:
1
Reps
2:
2
SE Channel
3:
1
Excite all reps w/Exchan 1
4:
24
Starting Freq. (units = 100 Hz)
5:
32
End Freq. (units = 100 Hz)
6: 500
No. of Cycles
7:
0
Rep Delay (units = 0.01 sec)
8:
2
Loc [ Pressure ]
9: -34.836
Mult
10:
0
Offset