SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES

7-16

CR510

12 V or 5 V

Ch E1

FIGURE 7.13-3. Hook up to AVW1

PROGRAM

AVW1 & CR510 USED TO MEASURE 1
GEOKON VIBRATING WIRE SENSOR.

* Table 1 Program

01:

60

Execution Interval (seconds)

01:

Excite-Delay (SE) (P4)
1:

1

Reps

2:

15

±

2500 mV Fast Range

3:

1

SE Channel

4:

1

Excite all reps w/Exchan 1

5:

1

Delay (units 0.01 sec)

6: 2500

mV Excitation

7:

1

Loc [ Temp ]

8:

.001

Mult

9:

0

Offset

02:

Polynomial (P55)
1:

1

Reps

2:

1

X Loc [ Temp ]

3:

1

F(X) Loc [ Temp ]

4: -104.78

C0

5: 378.11

C1

6: -611.59

C2

7: 544.27

C3

8: -240.91

C4

9:

43.089

C5

03:

Vibrating Wire (SE) (P28)
1:

1

Reps

2:

2

SE Channel

3:

1

Excite all reps w/Exchan 1

4:

24

Starting Freq. (units = 100 Hz)

5:

32

End Freq. (units = 100 Hz)

6: 500

No. of Cycles

7:

0

Rep Delay (units = 0.01 sec)

8:

2

Loc [ Pressure ]

9: -34.836

Mult

10:

0

Offset