5 vacuum system specifications, Vacuum system specifications -13, 8. vacuum system specifications -1 – Campbell Scientific PVS4100/4120/4150 Portable Automatic Samplers User Manual

Page 21

Advertising
background image

Section 1. Product Overview

1-13

1.3.5 Vacuum System Specifications

TABLE 1-8. Vacuum System Specifications

Feature Description

Switches

Run/Off (SPST Toggle).

Sample Volume

Adjustable, 50cc to 500cc [PVS4100 and
PVS4120]
Adjustable, 50cc to 250cc [PVS4150]

Sample Transport
Velocity

PVS4100: Minimum of 3 ft/sec at 20 ft of lift (3/8”
ID intake line).
PVS4100: Minimum of 3 ft/sec at 16 ft of lift (5/8”
ID intake line).
PVS4100: Maximum Vertical 27.5 ft.
PVS4120 and PVS4150: Minimum of 3 ft/sec at 13
ft of lift.
PVS4120 and PVS4150: Maximum Vertical: 18 ft
For complete charts, see page 1-15.

Metering Chamber

500cc capacity, clear acrylic, calibrated in 100cc’s.
500cc capacity, glass (Pyrex), calibrated in 100cc’s.
[OPTIONAL]
250cc capacity, clear acrylic, calibrated in 50cc’s.
[PVS4150]

Metering Chamber
Cover

Nylon
Reinforced Teflon Top. [OPTIONAL]

Volume Control Tube 316 Stainless Steel

Metering Chamber
Level Electrode

316 Stainless Steel

Intake Hose

Nylon-Reinforced PVC, 3/8” ID (standard 25 ft
with sinker)
Nylon-Reinforced PVC, 5/8” ID [OPTIONAL]
Teflon-lined Tygon, 1/2" ID [OPTIONAL]
Teflon-lined Tygon, 3/4" ID [OPTIONAL]

Discharge Hose

Latex, 3/8” ID
Latex, 5/8” ID [OPTIONAL]
Silicone, 3/8” ID [OPTIONAL]
Silicone, 5/8” ID [OPTIONAL]

Sinker/Strainer

Lead Sinker

Stainless Steel Sinker/Strainer [OPTIONAL]

Stainless Steel Sinker [OPTIONAL]