6 program type multicomposite, Program type - multi-composite -31, 6 program type - multi-composite – Campbell Scientific PVS4100/4120/4150 Portable Automatic Samplers User Manual

Page 71

Advertising
background image

Section 5. Programming

5-31

5.5.6 PROGRAM TYPE - Multi-Composite

The following sequence of entries are made
on the Touchpad to describe how the
Sampler controller is to store the samples it
takes, in the hardware specified in its
configuration. The ACTIVE SETTINGS are
not being altered.

1. Press the SET key.

2. Press the PROGRAM TYPE key.

3. Press an ARROW key. Continue until
Multi-Composite is shown on the display
(for storage in multiple container array).

4. Press the ENTER key.

The display will respond with the message
“MULTI-COMPOSITE Total Bottles nn”.
The two digits, nn, will be flashing,
indicating they can be changed by scrolling
with the arrow keys.

Press ENTER. The display now reads
nn Cycles per bottle”. A previously set
value is displayed. One of the digits is
flashing, prompting for a numeric entry.
Using the number keys (0-9), select the
number of times the same bottle will be used
at the times determined by the programming,
i.e., enter the number of samples to be taken
before the stepper advances to the next
bottle. After the first digit is entered, the
second digit will flash, prompting for the
second digit of the entry. Each sample
placed in a bottle will be a complete cycle.
The program will repeat this action each
time the sampling is initiated, until the
“Total Bottles” setting is reached.

Advertising