Reviewing the Test and Adjustment Results

Once the tests have finished running, you will want to review the results of the tests. The Agilent Service
Support Software window displays the DUT model number and serial number, the selected tests, the results
of the selected test, and the current test information. Six buttons, which are used to control the testing, are
also displayed.

The Selected Tests box displays the name of the selected tests, the pass/fail status (P/F) of each test that has
been run, the total number of points that each test checks, and the number of points that passed and failed for
each test. The pass/fail status indicates a failure if any point in that test fails.

The Selected Test Results box shows the results of the test that is highlighted in the Selected Tests box. The
Selected Test Results box shows the pass/fail status (P/F), the lower limits (LL), the measured value (Result),
the upper limits (UL), and the measured units (for example, kHz, mV, or dBm) for each test point checked by
the performance tests and some adjustments. “**” indicates values not displayed by the adjustments.

The Current test box shows the results of the test that is currently running. The Current test box shows the
pass/fail status (P/F), the lower limits (LL), the measured value (Result), the upper limits (UL), and the
measured units (for example, kHz, mV, or dBm) for each test point checked by the performance tests and
some adjustments. “**” indicates values not displayed by the adjustments.

The following buttons are also displayed on the Agilent Service Support Software window. These buttons are
used to control the testing. Only the buttons that are appropriate are active. For example, if the test is already
in progress, the Run button would not be active.

Run

Start running the highlighted test when initially starting the testing. Also used to start
testing at the same point where the test was stopped.

Stop

Stop the test that is currently running. The test stops after making the next measurement.

Restart

Rerun a test that was running when the testing was stopped. This restarts the test from the
beginning.

Next Test

Quit running the current test and give it a Fail status. Then, continue testing with the next
test.

Rerun

Restart the testing at the beginning of the first test.

Abort

Quit testing. Abort all tests.