Performance Tests
Dual Arbitrary Waveform Generator Check
3-40
out of phase and that they do not have any discontinuities. Refer to the illustration below.
• Frequency is approximately 64 kHz.
• Amplitude is approximately 1.9 V
p-p
.
• Both waveforms are symmetrical around 0 volts (the average voltage is approximately 0V).