Required Equipment

2- 3

Table 2-1

Required Equipment

Test Equipment

Preferred Model
or Alternate Model

Critical Specifications

Power Meter

Agilent E4418B E-Series
power meter,
or Agilent E4419A/B E-Series
power meter

The power meter and the
signal generator need to be
plugged into the same ac
power circuit. The power
meter also needs to be
connected to the signal
generator with a minimum
length (18") GPIB cable.

Power Sensor,
Input: Type-N (m)

Agilent E9304A E-Series
power sensor

The Agilent 8482A power
sensor can be used in place of
the Agilent E9304A E-Series
power sensor in some
performance tests and some
adjustments.

9 kHz to 6 GHz,

70 to +20 dBm

100 kHz to 4.2 GHz,

30 to +20 dBm

This power sensor is not an
alternate model because some
performance tests can not use
the Agilent 8482A power
sensor.

Spectrum Analyzer,
Input: Type-N (f)

or Input: 2.4 mm (m)

Option 001
Output: APC 3.5 (f)

Option 007
Output: APC 3.5 (f)

Option K35
Input: SMA (f),
Output: SMA (f)

Agilent 8563/64/65EC
Option 001 and K35
spectrum analyzer

or Agilent 8563/64/65E
Option 001, 007, and K35
spectrum analyzer

Option 001 adds a Second IF
Output

Option 007 adds digitized fast
time domain sweep; this
functionality has been
incorporated into the
Agilent 8563/64/65EC
spectrum analyzer.

Option K35 is required only
when adjusting

“WCDMA

Adjacent Channel Power
(Option 400)” on page 3-34

9 kHz to 6 GHz

WCDMA Test Set,
RF Input: Type-N (f),
LO Input: SMA (f),
IF Output: SMA (f)

Agilent Option K35 WCDMA
test set