Agilent Technologies E4438C User Manual

Page 44

Advertising
background image

Required Equipment

2-4

Function Generator,
Output: BNC (f)

Agilent 8904A function
generator

Function Generator,
Output: BNC (f)

Agilent 33120A function
generator

Option 001 is required only
when testing

“Digital

Modulation Level Accuracy”
on page 3-27

.

Two Required

Frequency Counter,
Channel 1: BNC (f)
Channel 2: BNC (f)
Channel 3: Type-N (f)

Agilent 53132A Option 050
frequency counter

Option 050
adds Channel 3 (5 GHz)

The frequency counter must
be plugged into the ac power
line for at least 24 hours
before testing begins.

Frequency Standard,
Output: BNC (f)

Use a 10 MHz house
standard. (For example, an
Agilent 5071A primary
frequency standard or
equivalent can be used.)

Vector Signal Analyzer,
Input: Type-N (f)
Output: BNC (f)

Agilent 89441A vector signal
analyzer with Options AYA,
AYH, AY7, AY9, B7A, and
UFG or UTH.

Option AYA adds vector
modulation analysis.

Option AYH adds digital
video modulation analysis.

Option AY7 adds second
10 MHz input channel.

Option AY9 adds extended
time capture memory
to 1 MSample.

Option B7A adds EDGE to
the vector modulation
analysis that is added by
Option AYA.

Option UFG adds extended
RAM and additional I/O.

Option UTH adds extended
RAM and additional I/O.

Table 2-1

Required Equipment

Test Equipment

Preferred Model
or Alternate Model

Critical Specifications