Performance Tests

Phase Modulation Accuracy and Distortion

3-12

Phase Modulation Accuracy and Distortion

These automated tests verify the phase modulation accuracy and distortion specifications. Accuracy is
verified directly with a measuring receiver, and distortion is verified by measuring the demodulated output
from the measuring receiver, using the audio analyzer. The audio analyzer provides a more accurate
verification with better resolution than using the measuring receiver alone.

Recommended Equipment

• Agilent 8902A measuring receiver

• Agilent 8903B audio analyzer with Options 010 and 051, or Options 011 and 050

Equipment Setup

Figure 3-10

Phase Modulation Accuracy and Distortion Setup