Analog output calibration, Specifications, Identification – NOVUS N1040i Indicator User Manual

Page 5: Safety information, Support and maintenance, Limited warranty and limitation of liability

Advertising
background image

N1040i Indicator

NOVUS AUTOMATION

5/6

INPUT CALIBRATION
All inputs are factory calibrated and recalibration should only be done

by qualified personnel. If you are not familiar with these procedures

do not attempt to calibrate this instrument.
The calibration steps are:
a) Configure the type of input to be calibrated.
b) Configure the lower and upper limits of indication for the

maximum span of the selected input type.

c) At the input terminals inject a signal corresponding to a known

indication value a little above the lower display limit.

d) Access the parameter “inL(”. With the keys

and

, f adjust

the display reading such as to match the applied signal. Then

press the

P

key.

e) Inject a signal that corresponds to a value a little lower than the

upper limit of indication.

f) Access the parameter “ink(”. With the keys

and

, adjust

the display reading such as to match the applied signal. Then

press the

P

key.

Note: When checking the indicator calibration with a Pt100 simulator,

pay attention to the simulator minimum excitation current

requirement, which may not be compatible with the 0.170 mA

excitation current provided by the indicator.
ANALOG OUTPUT CALIBRATION
1.
Configure analog input for 0-20 mA retransmission (

retr

=

p.0.20

).

2. Attach a milliamp meter to terminals 13 and 14.
3. Enter Calibration Cycle.
4. Access “

ovl(

” calibration parameter and press

followed by

.

5. Using

or

Keys, set “

ovLC

” parameter to the same value

read in the milliamp meter

6. Access “

ovk(

” calibration parameter and press

followed by

.

7. Using

or

Keys, set “

ovkC

” parameter to the same value

read in the milliamp meter.

8. Exit Calibration Cycle.

SPECIFICATIONS

DIMENSIONS: ............................................................ 48 x 48 x 80 mm

Approximate Weight: .............................................................. 75 g

POWER SUPPLY: ..........................100 to 240 Vac (±10 %), 50/60 Hz

Optional 24 V ...................... 12 a 24 Vdc / 24 Vac (-10 % / +20 %)

................................... 24 to 240 Vdc (±10 %) for model N1040i-F

Maximum consumption: ......................................................... 6 VA

ENVIRONMENTAL CONDITIONS:

Operation Temperature: ............................................... 0 to 50 ºC
Relative Humidity: .................................................... 80 % @ 30 ºC
For temperatures above 30 ºC, reduce 3 % per ºC
Indoor use; Installation Category II, Pollution Degree 2; altitude <

2000 meters

INPUT ................................................................ According to Table 01

Internal Resolution: ...................................... 32767 levels (15 bits)
Display Resolution: ................................................ 0.1 / 1 (°C / °F)
Input reading rate: ......................................... up to 55 per second
Precision @ 25 °C: .... J, K, T, E: 0.25 % of the span ±0.1 °C / °F

..................................... N, R, S, B: 0.25 % of the span ±1 °C / °F

................................................................ Pt100: 0.2 % of the span

..................... 4-20 mA, 0-50 mV, 0-5 V, 0-10 V: 0.2 % of the span

Input impedance: ........ Pt100, thermocouples, 0-50 mV: > 10 M

................................................................. 0-5 V, 0-10 V: > 500 k

.............................................................................. 4-20 mA: 100

Measuring of the Pt100: ......................... 3 wire type, (α=0.00385)
With compensation of the cable length, max 50 meters,

excitation current of 0.170 mA.

OUTPUT ALARM1: ..................... Relay SPDT; 240 Vac / 30 Vdc / 3 A
OUTPUT ALARM2: ............ Relay SPST-NA; 240 Vac / 30 Vdc / 1,5 A
RETRANSMISSION OF PV: .................................................................
.................................. 0-20 mA / 4-20 mA / 500

max. / 12.000 levels

24 Vdc SOURCE: ................................. 24 Vdc (±10 %) / 20 mA max.
CASE: .................................................. Polycarbonate (PC) UL94 V-2
BACK PANEL
: ....................................................... ABS+PC UL94 V-0
ELECTROMAGNETIC COMPATIBILITY: ............... EN 61326-1:1997

and EN 61326-1/A1:1998
SAFETY: .......................... EN61010-1:1993 and EN61010-1/A2:1995
ADEQUATE CONNECTIONS FOR TERMINALS OF THE CLAMP

TYPE;
STARTS OPERATION
: after 3 seconds connected to the power

supply;
CERTIFICATIONS
:

and

.

IDENTIFICATION

N1040i -

A -

B -

C

A: Outputs Features

RR

2 relays available (ALARM1 / ALARM2)

RA

1 relay and one analog output 0-20 / 4-20 mA

RE

1 relay and one auxiliary 24 Vdc voltage source

B: Digital Communication

485

RS485 available

C: Power Supply Features (POWER)

Blank 100 a 240 Vac/dc

24V

12 a 24 Vdc / 24 Vac

F

24 a 240 Vac/dc

SAFETY INFORMATION

Any control system design should take into account that any part of

the system has the potential to fail. This product is not a protection or

safety device and its alarms are not intended to protect against

product failures. Independent safety devices should be always

provided if personnel or property are at risk.
Product performance and specifications may be affected by its

environment and installation. It’s user’s responsibility to assure

proper grounding, shielding, cable routing and electrical noise

filtering, in accordance with local regulations, EMC standards and

good installation practices.

SUPPORT AND MAINTENANCE

This product contains no serviceable parts inside. Contact our local

distributor in case you need authorized service. For troubleshooting,

visit our FAQ at

www.novusautomation.com

.

LIMITED WARRANTY AND LIMITATION OF

LIABILITY

NOVUS warrants to the original purchaser that this product is free

from defects in material and workmanship under normal use and

service within one (1) year from the date of shipment from factory or

from its official sales channel to the original purchaser.
NOVUS liability under this warranty shall not in any case exceed the

cost of correcting defects in the product or of supplying replacement

product as herein provided and upon the expiration of the warranty

period all such liability shall terminate.
For complete information on warranty and liability limitations, check

appropriate section in our web site:

www.novusautomation.com

.

Advertising