7.1 TAKING CARE WITH THE SENSORS
The calibration of the humidity sensor may be changed if it is exposed to

contaminant vapors or extreme humidity and temperature conditions for long

time periods. For accelerating the calibration recovery follow the procedure

as described below:

• Remove the sensor from the capsule.
• If there are solid particles on the sensor, wash it with water.
• Place the sensor within a 80 °C (+/- 10°C) oven for 24 hours.
• Keep the sensor in a place at a temperature between 20 and 30 °C and

humidity over 75% RH for 48 hours.

• Place the sensor back in the capsule.

IMPORTANT

The sensor used in this transmitter may be damaged or become out of

calibration if it is exposed to chemical agents contaminated atmosphere.

Hydrochloric Acid, Nitric Acid, Sulphuric Acid and Ammonia in high

concentrations may damage the sensor. Acetone, Ethanol and Propylene

Glycol may cause a reversible measurement error.