NOVUS FieldLogger Register User Manual

Fieldlogger register, Introduction, Technical specifications

Advertising
background image

NOVUS AUTOMATION

1/8

FieldLogger Register

ELECTRONIC LOGGER – INSTRUCTIONS MANUAL V2.2x B

INTRODUCTION

FieldLogger is an equipment that reads and registers analog
variables. It works as an electronic data logger, storing logged
information in its internal memory for later analysis. It can also work
as a real-time remote measuring module, which does not have
internal memory and only logs information and transmits them to a
supervisory system.

Its 8 configurable input channels, digital input and alarm outputs,
backup power, etc, make it very versatile, which makes possible to
use it in a large number of applications.

It is configured through an easy-to-use software. The equipment is
provided with the software, which runs on Windows98 platform and
higher.

TECHNICAL SPECIFICATIONS

Signal input. ............................... Configurable according to Table 01.
Eight (8) channels: Ch1, Ch2, Ch3, Ch4, Ch5, Ch6, Ch7 e Ch8.
Note: All thermocouples are linearized through software and cold
junction compensation. NBR 12771 calibrated thermocouples, NBR
13773 Pt100.

Accuracy: .....................J, K, T, E, N: 0.2 % of the max. range ±1 °C;

.......................................R, S and B: 0.25 % of the max. range ±3 °C;
.............................................................Pt100: 0.2 % of the max. range
...................................4-20 mA and 0-50 mV: 0.2 % of the max. range

Input impedance: .................Thermocouples/Pt100/0-50 mV: >1 M

..........................................................................4-20 mA: 33

+ 1.5 V

PT100 measurements: 3-wire circuit with excitation current of 0.17
mA and cable-resistance compensation.
Note
: When using Pt100 calibrators to gauge the meter, check if the
required excitation current is compatible with the register current.

Internal resolution: ........................................ 20,000 levels (>14 bits)

Log capacity:.......................................................131072 (128 k) logs.

Sampling rate:..................From 550 ms to 950 ms for eight channels

Outputs (ALM1 and ALM2):.......SPST-NA 3 A / 250 V (3 A / 30 Vdc)

Power supply: .......................................100 to 240 Vac/dc, 50/60 Hz.

................................................................................Optional 24 Vac/dc
.......................................................................Max. Consumption: 2 VA

Secondary Supply (V

BAT

):........................Voltage: from +6 to 24 Vdc

..........................................................................Consumption: < 50 mA

Environmental Conditions:........ Operating Temperature

: 5 to 50 °C

Relative humidity: 80 to 31°C, linear fall of 50 to 40 °C; Internal
use; Installation Category II, Pollution Degree 2; altitude < 2,000 m

Dimensions: ......105

Ч

90

Ч

60 mm, Approximate weight: 210 grams

Protection:.................................................................................... IP30

Plastic ABS case

Standard serial communication RS485 MODBUS-RTU

INPUTS

The logger has 8 input channels. You can configure the type of signal
to be read by each channel. All types of input channels accepted are
factory calibrated and no further adjustment is required. The user
defines and configures the channels that will be in use, and disables
the others.

The types of input channels accepted by the logger and their
maximum ranges are listed in Table 01.

MEASUREMENT INTERVAL

The speed of input channels readings depends on the number of
channels enabled and type of input programmed for each channel.

Each channel requires a reading time of about 50 ms. Programmed
channels with Pt100 input need more than 50 ms for cable length
compensation calculations. 150 ms are required for internal
processing, regardless of the number of channels enabled. Thus, if
only one channel is enabled, readings will take place at every 200 ms
(50 + 150). If this channel is programmed to Pt100, 250 ms are
required between readings (50 + 50 + 150).
Required time to read the eight channels is 550 ms. If they are
configured as Pt100 input, required time will be 950 ms.

INPUT TYPE

MEASUREMENT RANGE

thermocouple J

-50.0 to 760.0

o

C (-58.0 to 1400.0

o

F)

thermocouple K

-90.0 to 1370.0

o

C (-130.0 to 2498.0

o

F)

thermocouple T

-100.0 to 400.0

o

C (-148.0 to 752.0

o

F)

thermocouple E

-35.0 to 720.0

o

C (-31.0 to 1328.0

o

F)

thermocouple N

-90.0 to 1300.0

o

C (-130.0 to 2372.0

o

F)

thermocouple R

0.0 to 1760.0

o

C (-32.0 to 3200.0

o

F)

thermocouple S

0.0 to 1760.0

o

C (-32.0 to 3200.0

o

F)

thermocouple B

150.0 to 1820.0

o

C (302.0 to 3308.0

o

F)

Pt100

-200.0 to 530.0

o

C (-328.0 to 986.0

o

F)

4-20 mA Lin. J

thermocouple J curve, -50.0 to 760.0

o

C

4-20 mA Lin. K

thermocouple K curve, -90.0 to 1370.0

o

C

4-20 mA Lin. T

thermocouple T curve, -100.0 to 400.0

o

C

4-20 mA Lin. E

thermocouple E curve, -35.0 to 720.0

o

C

4-20 mA Lin. N

thermocouple N curve, -90.0 to 1300.0

o

C

4-20 mA Lin. R

thermocouple R curve, 0.0 to 1760.0

o

C

4-20 mA Lin. S

thermocouple S curve, 0.0 to 1760.0

o

C

4-20 mA Lin. B

thermocouple B curve, 150.0 to 1820.0

o

C

4-20 mA Lin. Pt100

Pt100 curve, -200.0 to 530.0

o

C

4-20 mA Linear

Programmable indication range from -1999 to

9999

0-50 mV Linear

Programmable indication range from -1999 to

9999

Table 01 – Signals accepted by the logger

OUTPUTS

The logger has two relay outputs (ALM1 and ALM2) that can work as
Alarm or Digital Output.

When configured as Alarm, they act as configured by the user.

As Digital Outputs they are remotely controlled through serial
communication.

Advertising