Pipetting – BrandTech Scientific Transferpette S -8/-12 User Manual
Page 34
➠
➧
34
1.
Fitting the tips
Use the correct tips according to the volume
range or the color code. Ensure that the tips
are securely seated.
Pipette tips are disposables items!
Pipetting
2.
Volume setting
a) Push the volume-change protection upward
to disengage (UNLOCK).
b) Select the desired volume by rotating the
volume-setting wheel. Avoid twisting and
abrupt rotating motions during this adjust-
ment.
c) Push the volume-change protection down to
re-engage (LOCK). Note: The volume-change
protection tightens but does not lock volume-
setting wheel.
4.
Aspirate sample
a) Press pipetting button to the first stop.
b) Hold the pipette vertically and immerse
the tips into the liquid.
3.
Align the manifold
The manifold can turn freely in both directions.
Volume
range
Immersion
depth in mm
Waiting
time in s
0.1 µl - 1 µl
1 - 2
1
> 1 µl - 100 µl
2 - 3
1
> 100 µl - 1000 µl
2 - 4
1
> 1000 µl
3 - 6
3