Pipetting – BrandTech Scientific Transferpette S -8/-12 User Manual

Page 34

Advertising
background image

34

1.

Fitting the tips

Use the correct tips according to the volume
range or the color code. Ensure that the tips
are securely seated.
Pipette tips are disposables items!

Pipetting

2.

Volume setting

a) Push the volume-change protection upward

to disengage (UNLOCK).

b) Select the desired volume by rotating the

volume-setting wheel. Avoid twisting and
abrupt rotating motions during this adjust-
ment.

c) Push the volume-change protection down to

re-engage (LOCK). Note: The volume-change
protection tightens but does not lock volume-
setting wheel.

4.

Aspirate sample

a) Press pipetting button to the first stop.
b) Hold the pipette vertically and immerse

the tips into the liquid.

3.

Align the manifold

The manifold can turn freely in both directions.

Volume
range

Immersion
depth in mm

Waiting

time in s

0.1 µl - 1 µl

1 - 2

1

> 1 µl - 100 µl

2 - 3

1

> 100 µl - 1000 µl

2 - 4

1

> 1000 µl

3 - 6

3