Accuracy table – BrandTech Scientific Transferpette S -8/-12 User Manual

Page 39

Advertising
background image

English

39

Accuracy Table

Transferpette

®

S -8/-12

Final test values related to the nominal capacity (maximum
volume) or the indicated volume steps indicated on the instru-
ment, obtained when instrument and distilled water are equili-
brated at ambient temperature (20 °C/68 °F) and with smooth
operation. According to DIN EN ISO 8655.

!

20 °C

Ex

* A = Accuracy, CV = Coefficient of Variation

Volume range
µl

Volume step
µl

A*

± %

CV*

%

Increment
µl

Rec. type
of tips, µl

0.5 - 10

10

1.6

1.0

0.01

0.5 - 20

5

2

2

1

8

6

5 - 50

50

0.8

0.4

0.1

2 - 200

25

1.4

0.8

5

6

3

10 - 100

100

0.8

0.3

0.2

2 - 200

50

1.4

0.6

10

4

2

20 - 200

200

0.8

0.3

0.2

2 - 200

100

1.4

0.6

20

4

1.5

30 - 300

300

0.6

0.3

0.5

5 - 300

150

1.2

0.6

30

3

1.5