Accuracy Table
Transferpette
®
S -8/-12
Final test values related to the nominal capacity (maximum
volume) or the indicated volume steps indicated on the instru-
ment, obtained when instrument and distilled water are equili-
brated at ambient temperature (20 °C/68 °F) and with smooth
operation. According to DIN EN ISO 8655.
!
20 °C
Ex
* A = Accuracy, CV = Coefficient of Variation
Volume range
µl
Volume step
µl
A*
≤ ± %
CV*
≤ %
Increment
µl
Rec. type
of tips, µl
0.5 - 10
10
1.6
1.0
0.01
0.5 - 20
5
2
2
1
8
6
5 - 50
50
0.8
0.4
0.1
2 - 200
25
1.4
0.8
5
6
3
10 - 100
100
0.8
0.3
0.2
2 - 200
50
1.4
0.6
10
4
2
20 - 200
200
0.8
0.3
0.2
2 - 200
100
1.4
0.6
20
4
1.5
30 - 300
300
0.6
0.3
0.5
5 - 300
150
1.2
0.6
30
3
1.5
Advertising