Adjustment – easy calibration – BrandTech Scientific Transferpette S -8/-12 User Manual

Page 40

Advertising
background image

40

The instrument is permanently adjusted for aqueous solutions.
If the pipette operation is clearly inaccurate, or if the instrument must
be adjusted for solutions of different densities and viscosities or specially-
shaped pipette tips, adjustments can be made using the Easy Calibration
Technique.

1.

Check the volume, determine actual value

(see page 37).

2.

Remove the label window and the label.

Push the hook forward, raise it slightly and
then pull it back.

3.

Using a paperclip or an unused pipette tip,

remove the protective film (this protective film
can be discarded).

4.

Push the red adjustment slider completely

back, raise the volume-setting wheel (decou-
pling) and release the adjustment slider.

5.

Set the previously determined actual value

(adjustment value) with the volume-setting
wheel in the UNLOCK position.

A volume check is recommended after every
adjustment.

6.

Push the adjustment slider completely

back again, push the volume-setting wheel
downwards and release the adjustment slider.
Re-insert the label and the label window.

The change to the factory settings is indicated by the red adjustment
slider now visible in the label window.

Note:

Adjustment – Easy Calibration