Pipetting
5.
Discharge sample
a) Place the pipette tips against the wall of
the vessel. Hold the pipette at an angle
of 30-45° relative to the container wall.
b) Press the pipetting button slowly to
the first stop and hold it down.
For serum and liquids of high viscosity or
low surface tension, observe adequate
waiting time to improve accuracy.
c) The blow-out stroke empties the tips
completely: Press the pipetting button
down to the second stop.
d) While doing this, wipe the pipette tips
against the wall of the container.
e) Remove the pipette tips from the con-
tainer wall and let the pipetting button
slide back.
c) Let the pipetting button slide back
slowly. In order for the liquid to reach its
end position, leave the tips immersed
for a few seconds.