Commissioning, operation – WIKA D-11-7 User Manual

Page 13

Advertising
background image

13

WIKA operating instructions pressure transmitter model D-1x-7

2478159.05 12/2011 GB/D/F/E

GB

Types of sealing

Correct sealing of the process connections with parallel threads at the sealing face

must be made using

suitable flat gaskets, sealing rings or WIKA profile sealings.

The sealing of tapered threads (e.g. NPT threads) is made by providing the thread with additional

sealing material such as, for example, PTFE tape (EN 837-2).

For further information on seals see WIKA data sheet AC 09.08 or under www.wika.com.

6.2 Making the electrical connection

The pressure transmitter must be earthed via the process connection.

Use the instrument with shielded cable.

Earth the shield on at least one end of the lead.

Only ever use a power source that ensures a safe electrical isolation of the operating voltage in

accordance with IEC/DIN EN 60204-1. In addition, observe the general requirements for PELV

circuits in accordance with IEC/DIN EN 60204-1.

6. Commissioning, operation

Parallel threads

Tapered threads

per EN 837

per DIN 3852-E

NPT, R and PT

1

1

Advertising