Commissioning, operation – WIKA D-11-7 User Manual

Page 21

Advertising
background image

21

WIKA operating instructions pressure transmitter model D-1x-7

2478159.05 12/2011 GB/D/F/E

GB

Zero adjustment

Output data

Input data

Explanation

0x21

XX

XX

0x21

HI

LO

Read zero adjustment

0x23

HI

LO

0x23

HI

LO

Write zero adjustment

The respective correction value can be calculated as follows:

Correction value = -

Offset x 50,000

Span

Correction value: Digit value which is programmed into the instrument via service 0x23.

Offset: Deviation of the zero point from the set point in bar

Span: MBE - MBA in bar

Changes only take effect after a reset.

Zero adjustment, example 1:

A pressure transmitter with a measuring range of -1 ... +2.5 bar has a zero offset of +10 mbar, then one

would correct:
Measuring range -1 ... +2.5 bar

Span: 3.5 bar

Display value 10 mbar too high

Offset: +0.010 bar

6. Commissioning, operation

Correction value = -143

0xFF71

Correction value = - Offset x 50,000

Span

Correction value = - 0.010 bar x 50,000

3.5 bar

Output data

0x23

0xFF

0x71