Specifications, Gb 3. specifications – WIKA D-11-7 User Manual

Page 9

Advertising
background image

9

WIKA operating instructions pressure transmitter model D-1x-7

2478159.05 12/2011 GB/D/F/E

GB

3. Specifications

3. Specifications

Specifications

Models D-1x-7

Measuring ranges

bar

0.25 0.4 0.6 1

1.6

2.5

4

6

10

16

Overpressure limit

bar

2

2

4

5

10

10

17

35

35

80

Burst pressure

bar

2.4 2.4 4.8 6

12

12

20.5

42

42

96

Measuring ranges

bar

25

40

60

100 160

250

400

600

1,000

1)

Overpressure limit

bar

50

80

120 200 320

500

800

1,200 1,500

Burst pressure

bar

96

400 800 800 1,000 1,200 1,700

2)

2,400

2)

3,000

{Vacuum, gauge pressure, +/-, and absolute pressure are available}

Material
Wetted parts

(for other materials see WIKA diaphragm seals product range)

Model D-10-7

Stainless steel (with measuring range > 25 bar additionally Elgiloy

®

)

Model D-11-7

Stainless steel {Hastelloy}; O-ring: NBR {FPM/FKM or EPDM}

Case

Stainless steel

Internal transmission fluid

3)

Synthetic oil {halocarbon oil for oxygen applications}; {listed by FDA for

food industry}

Power supply U

B

DC

10 ... 30 V

Output signal

Profibus

®

DP protocol per EN 50170 / DIN 19245

Power consumption

W

1.7

Sensor services

2-byte error coding for sensor error or failure of electronics, min./max.

value upper deviation temperature + pressure

Terminating resistor

Internal terminating resistor can be activated via integrated DIP switch

Internal measuring rate

Hz

≤ 100

Warming-up period

min

< 10

Insulation voltage

DC

500 V

Accuracy

4)

% of span

≤ 0.25 {0.10} in the range 0 ... +50 °C

Non-linearity

% of span

≤ 0.04 (BFSL) per IEC 61298-2

Advertising