WIKA DPT-10 User Manual

Page 10

Advertising
background image

Density measurement

=

∆p

h g

+

h

Fig. 5: Density measurement with DPT10, h = defined mounting distance, Δp =
differential pressure, ρ = density of the medium, g = acceleration of gravity

1

DPT10

Interface measurement

1

+

h

2

3

Fig. 6: Interface measurement with DPT10

1

DPT10

2

Liquid with highest density

3

Liquid with lowest density

A metallic measuring cell is used as sensor element. The process
pressures are transmitted via separating diaphragms and filling oils to
the resistance measuring bridge (semi-conductor technology). The
differential-pressure dependent change of the bridge voltage is
measured, further processed and converted into a corresponding
output signal.

Functional principle

10

Differential pressure transmitter DPT10 - metal measuring diaphragm • Profibus PA

3 Product description

37245

-EN

-100120