6technical data – WIKA IUT-11 User Manual

Page 28

Advertising
background image

Universal IS Pressure Transmitter UniTrans

Technical Data

22

669

39

.0

5 D

/G

B

06

/2

00

6

Subject to change due to technical modifications.

© Copyright WIKA Alexander Wiegand GmbH & Co. KG / Germany

WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiegand-Str. · 63911 Klingenberg · (09372) 132 - 710 · Fax - 706 · E-mail: [email protected] · www.wika.de

103

6

Technical Data

6.1

Input-values

6.2

Output-values

Pressure Ranges
(Absolute pressure upon request)

/ overpressure limit / Burst pressure

0 ... 0.4 bar

2

2.4

0 ... 1,6 bar

10

12

0 ... 6 bar

35

42

0 ... 16 bar

80

96

0 ... 40 bar

80

400

0 ... 100 bar

200

800

0 ... 250 bar

500

1200

0 ... 600 bar

1200

2400

0 ... 1.000 bar

1500

3000

0 ... 1,600 bar

2000

4000

0 ... 2,500 bar

3000

5000

0 ... 4,000 bar

4400

7000

-1 ... +0 bar*

2

2.4

-1 ... +0,6 bar*

10

12

-1 ... +3 bar*

35

42

-1 ... +5 bar*

35

42

-1 ... +15 bar*

80

96

* only relative pressure
Do not exceed the nominal pressure!

Output signal

4 ... 20 mA, two-wire, optionally with
modulated HART signal

Accuracy [% of span]
(incl. non-linearity, hysteresis, non-
repeatability, zero signal and full
scale error)

< 0.10 at ranges < 1000 bar
< 0.30 at ranges > 1000 bar

Non-linearity

< 0.05% of span (BFSL) per IEC 61298-2
at ranges < 1000 bar
< 0.20% of span at ranges > 1000 bar

Turn down behavior: (1/k)
up to 1 : 5
1 : 5 to 1 : 20

no changes of deviation
the accuracy must be multiplied by the
factor (turn down / 5)
example for TD = 1:15, (k = 15)
accuracy = 0.10 * (15/5) = 0.3

Advertising
This manual is related to the following products: