WIKA IUT-11 User Manual

Page 49

Advertising
background image

22

669

39

.0

5 D

/G

B

06

/2

00

6

Universal IS Pressure Transmitter UniTrans

Operation of Transmitters with Display

Subject to change due to technical modifications.

© Copyright WIKA Alexander Wiegand GmbH & Co. KG / Germany

WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiegand-Str. · 63911 Klingenberg · (09372) 132 - 710 · Fax - 706 · E-mail: [email protected] · www.wika.de

124

9.5.2 Main Menu: Calibration of zero and span (with / without pressure)

A single pressure value is set for the zero point or the span end-point
within the sensor’s nominal pressure range, and assigned to the associ-
ated output current signal when making adjustments with existing pres-
sure. An error signal occurs when the existing pressure lies outside of
the sensor’s nominal pressure range. The value is not saved in this
case.

A mounting correction should be performed before or after making an
adjustment without pressure (dry adjustment) (see 8.4.3). The sensor
must therefore be placed in the reference position for the measurement
(installation site) without pressure on the diaphragm.

A mounting correction is unnecessary when making an adjustment with
pressure (wet adjustment). Otherwise, the mounting correction must be
performed before saving the zero point and span end- point.

A test / correction of the zero point is suggested after adjusting the
span in order to maintain optimum accuracy.

MAIN MENU

CALIBRATION

CALIBRATE
WITH PRESSURE

CALIBRATE
RANGE

CALIBRATE
SET ZERO

CALIBRATE
SET SPAN

CALIBRATE
WITHOUT
PRESSURE

CALIBRATE
DEFINE ZERO

CALIBRATE
DEFINE SPAN

DEFINE ZERO

DEFINE SPAN

Adjustment
Zero Point

Adjustment
Span

Adjustment zero point via entry
of a pressure value within the
nominal pressure range

Adjustment span via entry
of a pressure value within the
nominal pressure range

Important

Important

Advertising
This manual is related to the following products: