Mensor/WIKA Operating Instructions Series CPG 2500

39

Digital Pressure Gauge

SERIES CPG 2500

3.

Queries are special instructions in the form: [Command?] [Termination] where the
question mark, “?”, immediately precedes the terminator.

When a valid query is received, the CPG 2500 will return {data} terminated by CR and
LF. Floating point data is returned in the current engineering units in exponential format.

7.1.3.4 Output Formats

Pressure readings are returned in exponential notation in a format according to the
OUTFORM command as follows. Outform applies to both pressure channels.

1. <sp> pressure value <cr><lf>
2. <sp> pressure, units number,STANDBY <cr><lf>
3. <sp> pressure, pressure rate <cr><lf>
4. <sp> pressure, minimum peak, maximum peak <cr><lf>

7.1.3.5 Commands and Queries

Table 7.1.3.5 lists all of the current CPG 2500 commands and queries.

i

Notice

Channel specific commands are sent to only the active channel.
See ‘CHAN’ command.

Optional emulation modes are available in which a CPG 2500 can emulate remote func-
tions of different brands of pressure gauges. Please contact Mensor for more details.

Table 7.1.3.5 - CPG 2500 Commands and Queries

Command or
Query

Data

Response/Function

?

none

Returns data per the current output form.

ACQUIRE?

15 char string.
Example:
Acquire? Test_stand_1
Returns:
<sp>(yes or no),CCC...
CCC<cr><lf>

This command is used when multiple
computers would like to control the instru-
ment. Yes if acquisition is successful. No if
instrument is being controlled with another
computer. CCC... = name of controlling
computer. See: RELEASE? and UNLOCK

ADDRESS

0-31

Sets the GPIB Address

ADDRESS?

<sp> xx <cr><lf>

Returns the GPIB Address.