Mensor/WIKA Operating Instructions Series CPG 2500

55

Digital Pressure Gauge

SERIES CPG 2500

The ‘Q’ Command Data/Output Format

n

Description

Output Format

0

Pressure reading in the se-
lected units. This is the default
output format.

NNNNNNN<cr><lf>where each N is a number (0
through 9), +,-,decimal point, or a space. If the pres-
sure rate or peak monitor option is being used the
output format is NNNNNNN,NNNNNNN<cr><lf>. The
data before the comma is the pressure and the data
after the comma is the pressure rate or peak.

1

Raw A/D readings (for factory
use)

nnnnnnn,nnnnnnn<cr><lf> where each n is a number
(0 through 9) or a space.

2

Unit ID

Unit ID for the CPG 2500

3

Pressure range and measure-
ment units

NNNNNNN,NNNNNNN, @@@@@@@<cr><lf> where
each N is a number (0 through 9), decimal point, or a
space and each @ is an alpha-numeric character. The
output string represents the minimum pressure, maxi-
mum pressure and the pressure units.

4

Error status code

CPG 2500’s error codes

5

Calibration data

not emulated

6

Pressure rate or peak indica-
tion

nnnn where each n is a number (0 through 9) valid only
if optional pressure rate or peak is enabled. The mean-
ing of each digit is shown under the related feature in
the Options section of the DPG II manual.

7

Zero and span corrections

ZZZZZZZ,SSSSSSS<cr><lf> where ZZZZZZZ is the
zero correction and SSSSSSS is the span correction.

8

Calibration coefficients

not emulated

S Command: same as SPAN command.

T Command: used with RATE option; similar to DISPLAY command.