Digital pressure gauge series cpg 2500 – WIKA CPG2500 User Manual

Page 52

Advertising
background image

52

Mensor/WIKA Operating Instructions Series CPG 2500

Digital Pressure Gauge
SERIES CPG 2500

UNITS_TABLE

not emulated

UNITS_TABLE?

UNITS_TABLE? lf

not emulated

WINDOW

window sp
<value>lf

Sets the filter window for pressure readings where
<value> is a value in the current measurement
units within the range of the instrument.

WINDOW?

window?lf

Returns the filter window setting for the pressure
reading.
The instrument returns <value><cr><lf> where
<value> is a value in the current units within the
range of the instrument.

ZERO

zero sp <value>lf

This command requires that the PWZ or the PW
password must be issued before the first occur-
rence of the zero command. The zero command
sets the DPG zero offset to <value> in the current
units where <value> is a value between +/- 17 psi.

ZERO?

ZERO?lf

Returns the current zero calibration offset where
<value> is current zero value in current measure-
ment units.

7.1.3.8.2 UNITS Command Syntax for DPG 2100 Measurement Units

Table 7.1.3.8.2 - UNITS Command Syntax for DPG 2100 Measurement Units

N

Description

Output Format

Type

1

pounds per square inch

PSI

English

2

inches of mercury @ 0°C

INHG

English

3

inches of mercury @ 60°F

INHG

English

4

inches of water @ 4°C

INH2O

English

5

inches of water @ 20°C

INH2O

English

6

inches of water @ 60°F

INH2O

English

7

feet of water @ 4°C

FTH2O

English

8

feet of water @ 20°C

FTH2O

English