Mensor/WIKA Operating Instructions Series CPG 2500

Digital Pressure Gauge
SERIES CPG 2500

A?

<sp>n.nnnnne+nn<cr><lf>

Returns the A channel pressure reading.

AR?

<sp>n.nnnnne+nn<cr><lf>

Returns the A channel rate.

BARO?

<sp>n.nnnnne+nn<cr><lf>

Returns reading from barometric sensor

B?

<sp>n.nnnnne+nn<cr><lf>

Returns the B channel pressure reading.

BR?

<sp>n.nnnnne+nn<cr><lf>

Returns the B channel rate.

CALDISABLE

yes,no

Sets whether or not calibration of the ac-
tive channel is disabled.

CALDISABLE?

<sp>(yes or no)<cr><lf>

Returns whether or not calibration of the
active channel is disabled.

CERR

none

Clears the error queue.

CHAN

A or B

Sets the active channel on the instrument.

CHAN?

<sp>A or B or D or
Baro<cr><lf>

Returns which channel is active.

CHAN D

none

Sets the active channel to the Delta chan-
nel.

CHANFUNC

Press, peak, rate, baro

Sets the alternate function mode of the
active channel.

CHANFUNC?

<sp>CCCCC…<cr><lf>

Returns the alternate function mode of the
channel.

DECPT?

<sp>n<c><lf>

Returns the number of decimal points for
the active channel (see RESOLUTION).

DEFAULT

none

Sets the default values.

DELTAFUNC A-B

none

Sets the delta to be the result of the chan-
nel A reading - channel B reading.
(Chan D must be in the active channel).

DELTAFUNC B-A

none

Sets the delta to be the result of the chan-
nel B reading - channel A reading.
(Chan D must be in the active channel).

DELTAFUNC A+B

none

Sets the delta to be the result of the chan-
nel A reading + channel B reading.
(Chan D must be in the active channel).

DELTAFUNC?

<sp>{A-B or B-A}<cr><lf>

Returns delta function as one of the above
options. (Chan D must be in the active
channel).