Digital pressure gauge series cpg 2500 – WIKA CPG2500 User Manual

Page 43

Advertising
background image

Mensor/WIKA Operating Instructions Series CPG 2500

43

Digital Pressure Gauge

SERIES CPG 2500

RWINDOW?

<sp>n.nnnnne+nn<cr><lf>

Returns rate exponential filter window.

SBAUD

9600, 19200, 38400, 57600

Sets the serial baud rate.

SBAUD?

<sp>XXXX<cr><lf>

Returns the serial baud data.

SCREENSRC

A, B, or A,B, or B,A, or
Delta,A, or Delta,B

Sets the CPG 2500 display mode.

SCREENSRC?

<sp>{A,B or A,Delta or
Delta,B}<cr><lf>

Returns screen source as one of the
above options.

SDATA

7 or 8

Sets the serial data bits.

SDATA?

<sp>X<cr><lf>

Returns the serial data bits number.

SENSORID?

<sp>Mensor QRS,SN
XXXXXX,VER V.VV

Returns the active sensor’s serial number
and firmware version.

SPAN

desired pressure or ?

Sets span on active channel or for ?,
clears previous value, must be >50% FS
and has a 1% limit.

SPAN?

<sp>XXXXXXX<cr><lf>

Returns span scale factor for active chan-
nel.

SPARITY

even, odd, none

Sets the serial parity.

SPARITY?

<sp>CCCC<cr><lf>

Returns the serial parity.

SSTOP

1 or 2

Sets the serial stop bits.

SSTOP?

<sp>X<cr><lf>

Returns the serial stop bits.

UNITS

units code or text in Table
7.1.3.6.

Sets the engineering units for the active
channel.

UNITS?

<sp>CCCC<cr><lf>

Returns the pressure units of the active
channel in a text string.

UNLOCK

none

Releases Acquire locks. See ACQUIRE?
and RELEASE?

WINDOW

value in current units

Sets the exponential filter window for the
active channel.

WINDOW?

<sp>n.nnnnne+nn<cr><lf>

Returns the exponential filter window for
the active channel.

ZERO

desired pressure or ?

Sets zero to set pressure or for ?, clears
previous value.

ZERO?

<sp>xxxxxxx<cr><lf>

Returns zero offset for active channel.

Advertising