DOC

mm/dd/yyyy

Sets the date of cal for the active channel.

DOC?

<sp>mmddyy<cr><lf>

Returns the date of cal for the active chan-
nel.

DOM?

<sp>mm/dd/yyyy<cr><lf>

Returns the date of manufacture.

FILTER

Off, Low, Normal, High

Sets the reading filter 0, 80%, 92%, 95%.

FILTER?

<sp> (filter)<cr><lf>

Returns the reading filter.

GASDENSITY

Value in lb/cuft

Sets the head pressure gas density.

GASDENSITY?

<sp>n.nnnnne+nn<cr><lf>

Gets the head pressure gas density.

GASTEMP

Value in degrees F

Sets the head pressure gas temperature.

GASTEMP?

<sp>n.nnnnne+nn<cr><lf>

Gets the head pressure gas temperature.

GATEWAY

nnn.nnn.nnn.nnn

Sets the Ethernet gateway address.

GATEWAY?

<sp>nnn.nnn.nnn.nnn
<cr><lf>

Gets the Ethernet gateway address.

HEIGHT

Value in inches

Sets the head pressure height.

HEIGHT?

<sp>n.nnnnne+nn<cr><lf>

Gets the head pressure height.

ID?

<sp> MENSOR, 2500,
ssssss,v.v.vv

Ssssss is the serial number, v.v.vv is the
CPG 2500 software version.

IP

nnn.nnn.nnn.nnn

Sets the IP address of the instrument.

IP?

<sp>nnn.nnn.nnn.nnn
<cr><lf>

Returns the IP address of the instrument.

KEYLOCK

yes or no

Locks or unlocks keyboard.

KEYLOCK?

<sp>(yes or no)<cr><lf>

Returns yes or no.

LISTRANGE?

PRI,1, min,max

Returns the range of the installed sensor
for the active channel.

LOCALGRAVITY

value in ft/s^2

Sets the local gravity in feet/sec^2.

LOCALGRAVITY?

<sp>n.nnnnne+nn<cr><lf>

Returns the local gravity in feet/sec^2.

NETMASK

nn.nnn.nnn.nnn

Sets the Ethernet network mask.

NETMASK?

<sp>nnn.nnn.nnn.nnn
<cr><lf>

Gets the Ethernet network mask.

OUTFORM

1 to 4 - (see Table 7.1.3.6)

Sets the output format.

OUTFORM?

<sp>X<cr><lf>

Returns the output format (see 7.1.3.4).