2 reactive energy load performance, An365 – Cirrus Logic AN365 User Manual
Page 12
Advertising

AN365
12
AN365REV1
4.3.2 Reactive Energy Load Performance
Figure 11. Reactive Energy Load Performance
Power Factor
Load Current (A)
Current Dynamic Range (x:1)
Error
sin
= 1
120
1
-0.02%
60
2
-0.02%
12
10
-0.03%
1.2
100
-0.03%
0.24
500
0.03%
0.12
1000
0.01%
0.06
2000
0.02%
0.036
3333
0.09%
0.03
4000
0.08%
Lagging
= 0.5
120
1
0.19%
60
2
0.19%
12
10
0.19%
1.2
100
0.17%
0.24
500
0.20%
0.12
1000
0.26%
0.06
2000
0.23%
Leading
= 0.5
120
1
-0.23%
60
2
-0.22%
12
10
-0.24%
1.2
100
-0.24%
0.24
500
-0.2%
0.12
1000
-0.16%
0.06
2000
-0.17%
Table 10. Reactive Energy Load Performance
-1
-0.5
0
0.5
1
0.01
0.1
1
10
100
Percentage Error (%
)
Load Current (A)
sin
ࢥ = 1
Lagging sin
ࢥ = 0.5
Leading sin
ࢥ = 0.5