4.2.2 Reactive Energy Load Performance
Power Factor
Load Current (A)
Current Dynamic Range (x:1)
Error
sin
= 1
120
1
-0.01%
12
10
-0.03%
1.2
100
-0.04%
0.24
500
-0.05%
0.12
1000
-0.04%
0.06
2000
0.06%
0.036
3333
0.02%
0.03
4000
0.01%
Lagging sin
= 0.5
120
1
0.17%
12
10
0.24%
1.2
100
0.14%
0.24
500
0.10%
0.12
1000
0.07%
0.06
2000
0.07%
Leading sin
= 0.5
120
1
-0.20%
12
10
0.24%
1.2
100
-0.22%
0.24
500
-0.22%
0.12
1000
-0.22%
0.06
2000
-0.21%
Table 7. Reactive Energy Load Performance
-1
-0.5
0
0.5
1
0.01
0.1
1
10
100
Pe
rc
e
n
ta
ge
Er
ror (
%
)
Load Current (A)
sin
ࢥ = 1
Lagging sin
ࢥ = 0.5
Leading sin
ࢥ = 0.5
Figure 8. Reactive Energy Load Performance