5 restoring factory calibration, 6 storing a new factory calibration – Bacharach H25-IR User Manual

Page 33

Advertising
background image

H25-IR

Calibration

Instruction 3015-4342

27

3.5 Restoring Factory Calibration

NOTE: Factory calibration cannot be restored if the instrument was calibrated using an external leak source.

Restore an instrument to its factory calibration value as follows:

Display the Calibration Data menu by first pressing both the ENT and ESC buttons on either the instrument’s
front panel or Smart probe at the same time to display the Main menu. Next, use the Up and Down buttons to
highlight “CAL DATA”, and then press ENT to display the Calibration Data menu.

Main Menu:

SYSTEM

LOG
SETUPS
CAL

DATA

DIAGNOSTICS










From the Calibration Data menu, use the Up and Down buttons to highlight “RESTORE FACTORY CAL”, and
then press ENT to display the RESTORE FACTORY CAL screen.





Press the ENT button to confirm restoring the instrument to its factory calibration values.

3.6 Storing a New Factory Calibration


For all models fitted with an internal leak source, "factory" calibration data is stored in memory and used to
validate all subsequent internal calibrations performed by the user. An internal calibration failure will occur when
the calibration data collected during an internal calibration cycle, performed by the user, deviates from the
"factory" calibration data. This can occur when the internal particulate filter needs to be replaced or if a probe is
changed. When the internal filter or probe is replaced, or the optical sensor, pump or pneumatics is serviced, a
new "factory" calibration must be stored. This is accomplished by first performing an internal calibration, which
may fail due to the changes to the instrument, then pressing the ENT and CAL keys together.





Press ENT to confirm. This will take the data collected from the calibration just completed and store it as the
"factory" or reference calibration for validation of all future internal calibrations.

Calibration Data Menu:

CAL TYPE

EXT

RATE

EXT

GAS

TYPE

CAL

FACTOR

RESTORE FACTORY CAL

CAL INTERVAL

ENFORCE CAL INTERVAL

LAST CAL DATE

RESTORE FACTORY CAL
HIT ENTER TO CONFIRM

PRESS ENTER TO
STORE LAST CAL
AS REFERENCE

g/yr

FAULT

SETUP NO.1
SM SRCH

LAST CAL HAS
BEEN STORED
AS REFERENCE

g/yr

FAULT

SETUP NO.1
SM SRCH