H25-IR

Maintenance

Instruction 3015-4342

37

5 Maintenance

5.1 Diagnostics Menu

The Diagnostics menu provides information about the instrument’s operation that will be useful to service
technicians for the purpose of troubleshooting various fault conditions.

To display the Diagnostics menu, first press both the ENT and ESC buttons on either the instrument’s front
panel or Smart probe at the same time to display the Main menu. Next, use the Up and Down buttons to
highlight “DIAGNOSTICS” and then press ENT to display the Diagnostics menu.

Main Menu:

SYSTEM

LOG
SETUPS
CAL

DATA

DIAGNOSTICS







5.1.1 Sensor Data

The Sensor Data screen provides information about the sensor’s operating parameters and its surrounding
environment. When contacting the Bacharach Service Department, the operator may be asked to provide this
information to help in the troubleshooting process.

To display the Sensor Data screen, first display the Diagnostics menu as described in Section 5.1. Next, use
the Up and Down buttons to highlight “SENSOR DATA” and then press ENT.

Sensor Data Screen Information:

LR .......................... Measured Leak Rate
FAULT ................... Current Numerical Fault Code (refer to Section 5.1.2 Last Fault)
OPEN/CLOSED .... Measure / Search Flow Rate Valve Position
DET ....................... Detector Voltage
N ............................ Detector’s Output Noise Factor (with no gas detected)
Z ............................ Stored Zero Voltage of Detector
DET ....................... Detector Temperature
SPPM ..................... PPM Corrected to Standard Pressure and Temperature
BOX ....................... Enclosure Temperature
M ........................... Manifold Pressure
AMB ....................... Ambient Pressure
MPM ....................... Moles Per Minute
AU .......................... Average Optical Absorbency Unit of Gas being detected (0 = no gas)
PPM........................ Parts Per Million value of the Current Detected Gas Level
UML........................ Absolute Concentration of Gas in Micro Moles per Liter





Diagnostics Menu:

SENSOR DATA

LAST

FAULT

IR

EMITTER

PRESSURE
LOOP

TEST

CLR LAST FAULT

AUTO

GAIN