6 troubleshooting, H25-ir specifications – Bacharach H25-IR User Manual

Page 60

Advertising
background image

H25-IR

Specifications

Instruction 3015-4342

54

6 Troubleshooting

0001 BOX TEMP ERROR

Inside chassis

temperatures greater than

55degC or less

than -5degC

YES

NO

Move to cooler or warmer environment

to correct

Temp sensor defective

Replace motherboard

Does fault code

Also Indicate A/D failure

0400?

A/D defective

(other faults will also result)

Replace motherboard

YES

NO

Bacharach Inc. CONFIDENTIAL for internal use only

H25 IR FAULT CODE TROUBLE SHOOTER

0002 BENCH TEMP ERROR

Bench temp

sensor greater than

55degC or less

than -5degC

YES

NO

Move to cooler or warmer environment

to correct

Replace detector ribbon cable

And retest

Is detector ribbon

cable disconnected

on either end?

Reconnect cable and hot glue in place

YES

NO

Connect detector end

of a new bench with existing

cable. Is temperature

Normal?

Bad detector PCB

Replace optical sensor

YES

NO

Temperature reading

corrected with new

cable?

Bad ribbon cable, replace

Bad SPI Bus connection,

Replace motherboard

YES

NO

Bacharach Inc. CONFIDENTIAL for internal use only

H25 IR FAUL T CODE TROUBLE SHOOTER

Advertising