H25-ir operation – Bacharach H25-IR User Manual

Page 65

Advertising
background image

H25-IR

Operation

Instruction 3015-4342

59

4000 ZERO VOLT ERROR

Indicated the sensor voltage stored during

auto or manual zero was less that

3.9V or greater than 4.5V

Sensor voltage is zero

Or near zero

YES

NO

Detector cable loose

Or disconnected

NO

YES

IR Emitter resistance

Show **.* in diagnostic

Emitter screen

Emitter cable loose

Or disconnected

IR Emitter burned out, replace Optical sensor assembly, reset

digipot,re-cal box temp sensor and store new Factory calibration

Reconnect cable and readjust digipot

In factory menu

YES

NO

Reconnect cable and readjust digipot

In factory menu

Adjust sensor voltage with digipot in

Factory menu to 4.2V ±0.05V

YES

Bacharach Inc. CONFIDENTIAL for internal use only

H25 IR FAULT CODE TROUBLE SHOOTER

DETECTOR CABLE

CHECK BOTH ENDS

EMITTER CABLE

CHECK BOTH ENDS

8000 A/D CLIPPING ERROR

Indicates the sensor output

voltage is greater than 4.55V

Was new optical sensor

Just installed

YES

NO

Adjust sensor voltage with digipot in

Factory menu to 4.2V ±0.05V

Adjust Emitter power in diaganostic/

IR Emitter screen, per instructions

With new sensor

Adjust sensor voltage with digipot in

Factory menu to 4.2V ±0.05V

Store new “factory” cal

Bacharach Inc. CONFIDENTIAL for internal use only

H25 IR FAULT CODE TROUBLE SHOOTER