55

0004 MANIFOLD

PRESSURE SENSOR ERROR

Manifold pressure

Is greater than 16.0psia

or less than 3.0psia

YES

NO

Correct over or under pressure

Condition, check probe input and

exhaust output for pressure or

vacuum

Does fault code

Also Indicate A/D failure

0400?

A/D defective

(other faults will also result)

Replace motherboard

YES

NO

Snubber orifice blocked

Replace purple snubber

Stored atmospheric

pressure

Manifold pressure

Fault cleared?

NO

Bad Pressure transducer

Replace motherboard

Bacharach Inc. CONFIDENTIAL for internal use only

H25 IR FAULT CODE TROUBLE SHOOTER

0010 LOOP OPEN

Shorting jumper on

Current loop connector

Is missing or loop

resistance

>500ohms

YES

NO

Reconnect Jumper or reduce

Loop resistance

Retest with shorting jumper in place

of loop

Shorting jumper is

Connected?

A/D defective

(other faults will also result)

Replace motherboard

YES

NO

Does fault code

Also Indicate A/D failure

0400?

Loop circuit defective

replace motherboard

NO

YES

Bacharach Inc. CONFIDENTIAL for internal use only

H25 IR FAULT CODE TROUBLE SHOOTER