Commissioning, Switching on, Setting the safety limit values – IKA KS 3000 ic control User Manual

Page 16: Function shaking, Warning

Advertising
background image

16

Commissioning

Make sure before start-up of the equipment that the drain hose

leads into a drain!

For protection in transport, a strip of foamed material is inserted

between the shaker table and the inner face or the rear wall. This

must be removed before initial commissioning.

Check whether the voltage specified on the type plate matches the

mains voltage available. The power socket used must be earthed

(protective earth conductor contact). If these conditions are met,

Switching on

The appliance is switched on using the switch on the side of the

appliance. Once it has been switched on, all of the LEDs light up

briefly during the self test.

Main switch I/O

Display fields

Note the load guideline values for the

attachments prior to switching appliance

on (see „Attachments“).

Tapping the Start/Stop keys or On/Off key starts the particular

functions.

The desired default values can be changed using the + or - keys.
Use the shift key  to switch from hour / minute mode to minu-

te / second mode.

If the elapsed time exceeds the value of 100 hours, the display
switches from hour / minute mode

to hour mode

Only whole hours are displayed in hour mode.
The following sections contain detailed information about the
individual functions.

Setting the safety limit values

If the Time ON/OFF key  is held down while „SAFE“ is displayed, the safety limit values can be changed using the respective
up/down keys.

Function Shaking

Press  key to start or stop the shake function. The speed

can be adjusted during operation. The displayed value flashes

until the pre-set speed has been reached.

When the shaking function is started using button  , the timer

automatically starts measuring the time until the next switch-off.

Note:

The shaking function can only be started when the cover is

closed. When the cover is opened, the shaker automatically

switches off the shaking and heating functions.

the device is ready to operate after plugging in the mains plug.

If these procedures are not followed, safe operation cannot be

guaranteed and/or the equipment may be damaged.

Observe the ambient conditions (temperature, humidity, etc.) listed

under Technical Data.

Operating mode

Software

version

max. ad-

justable

temperature

max. ad-

justable

speed

Safety limit

values

WARNING