Timer, Function heating, Operating modes – IKA KS 3000 ic control User Manual

Page 17: Setting the operating mode, External temperature sensor, Warning

Advertising
background image

17

Timer

The desired shaking duration is set on the operator panel of the timer using the Time +/- keys. The shift key can be used to switch from
hour mode to minute mode.

The Time On/Off key activates the timer and shaking functions.

When the time is up, the shaking function stops and an acoustic signal is given off. If the cover is opened before the time is up, the
timer goes to „pause“ and the display flashes. If the cover remains open for more than 15 minutes, the timer switches off and an error
code appears on the display. The shaker no longer starts when the cover is closed.

If the timer is not activated, the Time display automatically shows the operating time following the start of the shaking function.

Function Heating

Once the appliance starts, the display indicates the actual value for the set temperature.

Press the  key to start or stop the heating function. The temperature is entered using the Temp +/- keys. In operation, the
temperature is displayed in 0.1 °C steps.

The target value appears approx. every 5 seconds and remains on the display for 2 seconds.

Note:
The heating function can only be started when the cover is closed. When the cover is opened, the shaker automatically switches off
the shaking and heating functions

Operating modes

You can choose between operating modes , and .

Appliance does not start up again following power outage Safety limit value for speed and temperature adjustable

Appliance starts up again following power outage Safety limit value for speed and temperature adjustable

Appliance starts up again following power outage Safety limit values cannot be changed

Setting the operating mode

Switch on appliance and simultaneously hold down the Time Start/Stop key  and the shift key  .

The display indicates either , or depending on the operating mode set (factory default setting A).
Switching the appliance off and on switches to the next operating mode.

Following the selection of the operating mode, the corresponding letter appears on the display for 5 seconds, the appliance is ready
for operation..

External temperature sensor

If external temperature sensor PT1000.80 is connected to the internal slide-on receptacle, the temperature can be measured at any
place in the medium.

Connector for

PT1000.80

The temperature control of the heating works automatically with this display

or measuring value. This is indicated by the lighting up of the LED next to the

temperature sensor symbol.

Locate the external tempe-

rature sensor and the

connecting cable so that

the shaking motion is unimpeded and no vessels
are damaged or tipped over.

WARNING