Reset calibration - temperature, Appliance variant ks 3000 ic control – IKA KS 3000 ic control User Manual

Page 18

Advertising
background image

18

To restore the factory settings, hold down the Mot „-“ key and the Temp „-“ key and press the main switch I/O.

Factory setting:-

Operating mode A

Upper

speed

limit

500

rpm

Safety

temperature

limit

95

°C

Temperature

calibration

Reset

Calibration - Temperature

Calibration with or without inserted sensor
- Fill Erlenmeyer flask (250ml) with water to the 100ml mark
- Place the Erlenmeyer flask in the centre
- Immerse the sensor of the external measuring device in the water
- Set the target temperature
- Close the cover and start the Temp function
- Wait until the temperature in the incubation cover has stabilised
- While holding down the shift key , briefly press the Temp On/Off key  , the display appears;
keep holding down the shift key .
- Set the temperature value read off the external measuring device using the Temp up + / down - keys
- Calibration is complete when you release the shift key

The appliance is calibrated at the factory.

This function allows the temperature to be calibrated to a desired value. This may be necessary, for example, if special sample
containers, their arrangement or external influences (e.g. sunshine) cause the measuring value for the temperature to be incorrect.

Appliance variant KS 3000 ic control

There is a cooler built into this appliance variant.

By using an external cooling unit, the working temperature can be lowered in relation to the room temperature (depending on the
supply temperature). Set the desired temperature on the KS 3000 and switch the heating function on.

The cooler is connected to an external cooling unit (e.g. IKA KV 600) via the plug connection at the back of the appliance. The
inlet and outlet connections are labelled accordingly at the back of the appliance. To connect the cooling pipes there are two hose
connection pieces included. They can be connected to a hose with a 10mm interior diameter. The connectors are unlocked by coaxially
positioning the unlocking lever and pressing slightly in the direction of the arrow. By positioning and light pressure in the direction of
the plug, the connectors are locked/connected to the inlet/outlet plug connections on the housing.

Drain connection

Inlet connection

Observe the maximum permissible pressure of 1 bar! As a safeguard, we recommend using a pres-
sure limiter (e.g. IKA C25). This is not necessary when using an IKA KV600.
Accumulated condensate is drained out of the cooler through the drain hose.

Water is the only coolant that

may be used (with antifreeze,

e.g. ethylene glycol).

Permissible cooling agents - inlet

temperature 3 to 20 °C
Temperature limit 80°C

Advertising
This manual is related to the following products: