Error codes

An error is indicated by an acoustic signal and an error code is displayed.

E. g.:

If faults cannot be eliminated directly, you must perform a RESET (see „Resetting parameters to the factory settings“)!

If the faults still cannot be eliminated, the appliance must be inspected by a technical service.

Error Description

Cause

Effect

Solution

code
Er 3 Temperature inside unit too high

• Permitted ambient temperature has

Heating off

• Switch off the unit. Allow it to cool down

been exeeded

and switch on again.

• Ventilation slots or fan housing

• Clean ventilation slots or fan

blocked

housing

• Observe maximum permissible ambient-

temperature

Er 4 Differance between setpoint and

• Motor seized or overload

Motor off

• Reduce the torque load

actual speeds is too large

(load)

• Reduce setpoint speed

Er 8 The calibration value of the tempe- • Fault in calibration procedure

Heating off

• Repeat the calibration procedure

rature sensor is outside the limit

• Value was incorrectly stored to memory

value

• EPROM switching error

Er 14 External temperature sensor,

• Short-circuit at temperature sensor

Heating off

• Check the plug

short-circuit

plug

• Substitude the temperature sensor

• Short-circuit in connecting cable

or on Temperature sensor

Er 16 External temperature sensor has

• SAFE Temp is set at a temperature

Heating off

• Allow the unit to cool down

exceeded the SAFE Temp

lower than the actual temperature on

• Set the SAFE Temp at a higher temperature

the external temperature sensor

Er 17 Temperature sensor - Incubation-

• SAFE Temp is set at a temperature

Heating off

• Allow the unit to cool down

room has exceeded the

lower than the actual temperature on

• Set the SAFE Temp at a higher temperature

SAFE Temp

the external temperature sensor

Er 26 Difference between the internal

• Ventilation slots in incubation room

Heating off

• Switch off the unit. Allow it to cool down

control and safety temperature

blocked

and then switch on again.

sensors too large

• Radial-flow fan does not rotate

• Check fan or ventilation slots and

• Fault in the control or safety

clean if necessary

temperature sensors

Er 60 Power outage

• Power outage during operation

Interruption of heating

• Delete the display by pressing the

or shaking function

shift key

PC 1 In remote operation (PC) with

•PC does not send data during the

Heating off

• Change the watchdog time

watchdog function 1 (WD) enabled: watchdog time

Motor off

• Send data from the PC within the

No communication between

• The connection/cable to the PC is

watchdog time

PC and unit

broken

• Check the connector cable and plug

PC 2 In remote operation (PC) with

• PC does not send data during the

The setpoint temperature • Change the watchdog time

watchdog function 2 (WD) enabled: watchdog time

is set to the WD safety

• Send data from the PC within the

No communication between

temperature

watchdog time

PC and unit

• The connection/cable to the PC is

The setpoint speed

• Check the connector cable and plug

broken

is set to the WD safety

speed