Communication between laboratory instrument and pc – IKA KS 130 control User Manual

Page 32

Advertising
background image

RMP_IN_X

X=4

Reading the real segment
number of ramp. With ramp
not started: 0

RMP_IN_X_y

X=4

Reading the accumulated value
and the ramp segment duration
(hh:mm:ss) of ramp segment y.

RMP_OUT_y n

hh:mm:ss

X=4

Setting the accumulated value (n)
and the ramp segment duration
(hh:mm:ss) for ramp segment y

RMP_START_X

X=4

Starting the ramp function, begin-
ning with ramp segment No. 1.
(Only possible after prior
START_X.) After RMP_STOP_X
START_X is not necessary.

RMP_STOP_X

X=4

Switching off ramp function.
Rated value=0 (Ramp is main-
tained, that means , ramp can be
restarted with RMP_START_X.

RMP_PAUSE_X

X=4

Stopping the ramp function.
Freezing of real rated value and
real ramp segment time

RMP_CONT_X

X=4

Continuation of ramp function
(After prior RMP_PAUSE_X).

RMP_RESET_X

Switching off ramp functions
and deleting of all set ramp
segments

RMP_LOOP_SET_X

X=4

To work off the ramps in one
loop

RMP_LOOP_RESET_X X=4

Ending of ramp loop

IN_TYPE

Requiring identification of
laboratory instrument

IN_NAME

Requiring the designation name

OUT_NAME

name

Output of designation name
(max.6 characters, default: IKA_S_)

OUT_WD1@m

Watchdog mode 1: If the WD1
result occurs, the agitation pro-
cess is turned off and Er2 is dis-
played. The watchdog time is set

to m (10...1800) sec with echo of
the watchdog time.
This command starts the watch-
dog function and must always be
sent within the time set by the
watchdog.

OUT_WD2@m

Watchdog mode 2: If the WD2
result occurs, the target speed
is set to the WD target safety
speed that has already been set.
The warning WD is displayed.
The WD2 result can be reset
with OUT_WD2@0.
The watchdog time is set to m
(10...18000) sec with echo of
the watchdog time. This com-
mand starts the watchdog func-
tion and must always be sent
within the time set by the
watchdog.

OUT_SP_42@n

Sets the WD target speed with WD
echo of the value set.

(KS 130 control version only)

PC 5.1 Adaptor IKA - Control
The Adaptor PC 5.1 splits the analog and serial signals. The ana-
log output signals are alotted to a 7-polar coupling according to
the Namur-recommendation; the serial signals to a 9-polar Sub-
D-bushing plug (RS 232 C).

PC 1.5 Cable
This cable is required to connect the 7-pin connector to a PC (25-
pin connector)

32

032007

Communication between laboratory
instrument and PC