Attachments supported weight (load) – IKA KS 260 control User Manual

Page 28

Advertising
background image

28

HS/KS260 0305

Attachments recommended by IKA must be connected to the
agitation table to ensure safe and secure operation. in addition ,
the usable weight and speed ranges shown in the diagrams (hat-
ched areas) must not be exceeded.

AS 260.1 Universal attachment
Universal, steplessly adjustable clamping cylinders make it posi-
ble to adjust to any container shape.

The universal attachment consists of:
AS 1.60 basic suspension

1x

Dimension WxHxD

AS 1.5 fastening screw

8x

in mm

AS 1.61 clamping cylinder

4x

425 x 135 x 334

ATTENTION: The base suspension of the universal attachment
is secured in place by four lateral clamping screwson the agitati-
on table.

AS 260.2 Holding bracket adapter
With additional holding brackets (AS2.1, AS2.2, AS2.3, AS2.4
and AS2.5) the holding bracket adapter is suitable foe working
with round flasks, measuring flasks and Erlenmeyer flasks.

Recommended / maximum fitting
AS 2.1 Holding bracket

12 / 12

Dimension WxHxD

AS 2.2 Holding bracket

8 / 9

in mm

AS 2.3 Holding bracket

5 / 8

330 x 24 x 334

AS 2.4 Holding bracket

4 / 4

AS 2.5 Holding bracket

2 / 4

ATTENTION! The holding bracket adapter is secured in place
by four lateral clamping screws on the agitation table.

AS 260.3 Dish attachment
For gentle agitation at low speed, for example for Petri dishes or
culture flasks. The dish attachments is equipped with an anti-slip
sheet that prevents the dish from moving around during the agi-
tation process.0

Dimension WxHxD
in mm
410 x 33 x 370

ATTENTION! The dish attachment is secured in place by four
lateral clamping screws on the agitation table.

AS 260.5 Separating funnel attachment
For intensive agitation of sparating funnels.

Max. fitting:

Dimension WxHxD

6 x 50 ml Separating funnel

in mm

5 x 100 ml Separating funnel

228 x 95 x 234

3 x 250 ml Separating funnel
3 x 500 ml Separating funnel

ATTENTION! The separating funnel attachment is secured in
place by four lateral clamping screws on the agitation table.

In order to ensure safe and secure operation, the agitator must
only be operated within the range of usable weight and speed
(see diagram - hatched area). Please make certain that the hol-
ding surface on which the agitator is placed is clean and level.
Otherwise the working range shown in the diagram (the hatched
area) cannot be used.

If the agitator is screwed together with an adjacent piece of
equipment (i.e. if it is integrated into a system) it should be noted
during operation that dynamic forces may arise as a result of
unfavorable loading circumstances and the position of the center
of gravity. This may cause the adjacent equipment to vibrate or
may cause the table to vibrate uncontrollably. If you notice that
the device is running unevenly, the speed must in any case be
reduced until the running irregularities no longer occur.
Make certain that individual agitation vessels are placed in the
middle of the agitation table and multiple agitation vessels are

Attachments

Supported weight (load)