29

place evenly on the agitation table, and that all vessels are secu-
rely fastened in place.

AS 260.1 Universal attachment

AS 260.5 Separating funnel attachment

Basic error codes:

ATTENTION! If an error code is encountered, an audio warning
signal is generated and the red Off signal light goes on in the dis-
play field. First, please try to continue operation by turning the
device off and back on again.If it is still not possible to eliminate
an error after an extended wait, please contact our service
department. If you do so, you should always tell us what error
code has been encountered. This simplifies the process of

tracking down the error and makes it possible to form a prelimi-
nary opinion.

Error code

Error

Cause of the error

Off - LED (red)

The potentiometer for

-Too rapid a change in the

1x flash /

the speed is not assig-

speed target value

2 sec pause

ning any target value

-Internal device error

Off - LED (red)

Permissible motor tem- -Permissible supported

2x flash /

peratur exceeded

weight exceeded

2 sec pause

-Permissible ambient
temperature exceeded

Motor locked

-Agitation table is being
hindered in this stroke mo-
tion by an outside effect

Read fork signal not

-Internal device error

available

Off - LED (red)

Maximum adjustable

-Resonance behavior of

3x flash /

speed of the device

the experiment setup

2 sec pause

exceeded

Off - LED (red)

Motor standstill for de-

-When turning on the

4x flash /

vice switch-on proce-

device, you must ensure

2 sec pause

dure

(safety confirma-

that the motor is not

tion not ensured

performing any rational
motion

Off - LED (red)

Safety relay cannot

-Internal device error

5x flash /

be activated

2 sec pause

Off - LED (red)

The switch for the TIME -Internal device error

6x flash /

function is not assigning

2 sec pause

any target value

Off - LED (red)

Problem with the power -The device is beeing

7x flash /

supply

operating with under vol-

2 sec pause

tage or overvoltage (for
permissible voltage range,
see technical data)

HS 260

KS 260

1/min

g

0

100

150

200

250

300

350

400

450

500

50

0

8000

7000

6000

5000

4000

3000

2000

1000

g

1/min

HS 260

KS 260

0

100

150

200

250

300

350

400

450

500

50

0

8000

7000

6000

5000

4000

3000

2000

1000