IKA KS 260 control User Manual

Page 31

Advertising
background image

31

HS/KS260 0305

• RTS:

(Pin 7) LOW (positive tension)
/ PC may send

• RTS:

(Pin 7) HIGH (negative tension)
/ PC may not send

• CTS:

(Pin 8) LOW (positive tension)
/ PC recipient

• CTS:

(PIN 8) HIGH (negative tension)
/ PC not recipient

• Access method: Data communication from laboratory instrument to

processor is only possible on demand of the processor.

Instruction Syntax
Here applies the following:
• The instructions are generally sent from the processor (master) to the
laboratory instrument (slave).
• The laboratory instrument exclusively sends on demand of the proces-
sor. Even error codes cannot be spontaneously communicated from the
laboratory instrument to the processor (automatic system).
• Instructions and parameters as well as subsequent parameters are sepa-
rated by at least one blank. (Code: hex 0x20)
• Each individual instruction including parameters and data as well as each
reply are terminated with CR LF (Code: hex 0x0D and 0x0A) and have a
maximum lenght of 80 characters.
• The decimal separator in a floating point number is the point (Code: hex
0x2E).
The above statements largely correspond with the recommendations of the
NAMUR-Association. (NAMUR-recommendations for the desing of electric
plug connections for the analog and digital signaltransmission to laboratory-
MSR individual units. Rev. 1.1)
Overview of the NAMUR-Instructions
(KS 130 control version only)
Abbreviations:
X,y =

numbering parameter (integer number)

M

=

value of variable, integer number

n

=

value of variable, floating point number

X

= 4

speed

X

= 6

speed range limit („safety“ speed)

NAMUR Instructions

Function

Display

additional

IN_PV_X

X=4

Reading the real-value

OUT_SP_Xn

X=4

Setting the actual value to n
up to a maximum of the set
upper speed limit

IN_SP_X

X=4; 6 Reading the set rated value

START_X

X=4

Starting the instrument’s

Remote

(remote) function

STOP_X

X=4

Switching off the instrument

Remote

function. Variables set with
OUT_SP_X are maintained. Con-
tains the instruction MP_STOP

RESET

Switching off the instrument
function

STATUS

Display of status
1*: A mode of operation
2*: B mode of operation
3*: C mode of operation
*0: manual operation without

fault

*1: automatic operation Start

(without fault)

*2: automatic operation Stop

(without fault)

<0: error code: (-1)

-2: Er2 / WD
-3: Er3
-4: Er4
-9: Er9
-41: Er41
-42: Er42
-83: wrong parity
-84: unknown instruction
-85: wrong instruction

sequence

-86: invalid rated value