Interface and outputs – IKA KS 260 control User Manual

Page 30

Advertising
background image

30

HS/KS260 0305

Off - LED (red)

Triac cannot be activated -Internal device error

8x flash /

2 sec pause

Control error codes:

ATTENTION! If an error code is encountered, an audio warning
signal is generated and the red Off signal light goes on in the dis-
play field. First, please try to continue operating by turning the
device off and back on again. If it is still not possible to elimina-
te an errorafter an extended wait, please contact our service
department. If you do so, you should always tell us what error
code has been encountered. This simlifies the process of
tracking down the error and makes it possible to form a prelimi-
nary opinion.

Error code

Error

Cause of the error

Er 2

No communication bet- -Interface not connected
ween PC and control de- -PC is not sending any
vice in remote operation data within the amount
with active watchdog func- of time set by the watch-
tion in Mode 1

dog

Er 3

Internal device tempe-

-Permissible ambient

rature too high

temperature exceeded

Er 4

Motor locked or over-

-Agitator table is being

loaded, Problem with

hindered in its stroke motion

read fork signal

-Internal device error

Er 9

Error while reading

-BLP Logik

stored values.

Er 41

Triac defective

-Internal device error

Er 42

Safety relay defective

-Internal device error

WD

No communication bet- -Interface not connected
ween PC and contol de- -PC is not sending any
vice in remote operation data within the amount
with active watchdog func- of time set by the watch-
tion in Mode 1

dog

(HS/KS 260 control version only)

The HS/KS 260 control is equipped with a 15-pin SUB-D connector on the
rear side of the device. The pins are assigned in operatin with analog and
serial signals.

Analog autput
Voltage values for the speed measurement quantify
are present on the analog pins.

(10)

Analog GND

(15)

Speed measurement value

100mVDC / 100 1/min

RS 232 C serial interface
The serial assignment of the socket can be used to control the device exter-
nally by means of a PC and a suitable
applikation program.

Configuration of the serial RS 232 C interface

• The function of the interface line between the labora-
tory device and the automation system is a selection of
the signals specified in EIA Standard RS 232 C, corre-
sponding to DIN 66020 Part 1. For the assignment of the
signals, please refer to the illustration.
• Standard RS 232 C applies to the elctronic proerties of
the interfaces and the assignment of signal states in
accordance with DIN 66259 Part1.
• Transmission procedure: Asynchronous character

transmission in start-stop
mode

• Type of transmission:

full duplex

• Character format::

Character creation according to the data format
in DIN 66022 for start-stop mode. 1start bit;
7character bits; 1parity bit (even); 1stopbit.

• Transmission speed:

9600 baud

• Data flow conrol:

RTS/CTS hardware handshake

Interface and outputs

5

1

0

1

5

1

6

1

1

Analog GND

100mV = 100 1/min

5

1

0

1

5

1

6

1

1

Analog GND

RS232_GND

TxD

RxD

RTS

CTS

100mV = 100 1/min