Caron CRSY102 User Manual
Page 11
![background image](/manuals/556066/11/background.png)
CRSY102 Operations Manual
Rev G 1/31/2013
Page 11
Recirculator
Inlet
Recirculator
Outlet
Power
Cord
SECTION 2 -- INSTALLATION
Choosing a Location
Locate recirculator near the corresponding chamber(s) so recirculator inlet is
physically lower than chamber drain(s). This should be a dry, clean, and level area.
Locate the unit in an area out of direct sunlight and where the water level viewing
window can easily be seen. The unit is on casters for easy mobility. Use the green
front and black rear handles to lift the recirculator.
Connecting to Chamber Drain (Recirculator Inlet)
Connect the chamber drain(s) to the recirculator inlet (
3/8” OD, larger fitting
connection) using the blue tubing and fittings provided. Cut tubing to proper length
to ensure the shortest run possible while allowing sufficient tubing slack so the
recirculator can be moved (if necessary) and side panels accessed.
Chamber(s) height may need to be raised for proper drainage.
The unit sits low to the ground and can bottom
out while moving over bumps.
Because the chamber drain / recirculator inlet
(blue) line is a gravity drain, there can
’t be any
kinks, traps, or vertical rises in the line.