DI Cartridge

O-ring

Filter
Housing

6. Discard used DI cartridge and insert new DI cartridge into tall filter housing.

7. Apply silicone grease to the o-ring.
8. Using the filter wrench, screw tall filter housing back in place until tight (tighten

1/8 turn after hand tight).

9. Using the filter wrench, unscrew the short filter housing.

10. Discard used carbon filter & insert new carbon filter into short filter housing.





11. Apply silicone grease to the o-ring.
12. Using the filter wrench, screw short filter housing back in place until tight.

(tighten 1/8 turn after hand tight).

13. Install right and left side access panels. Place the panel bottom into each

groove and rotate panels upward until latches catch.

14. If necessary, roll recirculator back into place. Do not kink any tubing lines.
15. Plug power cord in and turn recirculator on.

Be sure o-ring inside of the filter housing is
properly seated.